Hirayama HICLAVE HVP-50 Floor-Standing Autoclave with Heat/Vacuum Drying
Advanced Laboratory Sterilization & Drying Solution - 50L Capacity, 120V
Technical Specifications
-
Chamber Volume: 50L
-
Sterilization Temperature Range: 105°C to 135°C
-
Chamber Dimensions: 11.8" × 28" (30 × 71 cm)
Advanced Features & Performance Benefits
-
Superior Drying Capability: Fully automatic heat/vacuum drying cycle delivers faster and more efficient drying for solid loads
-
Enhanced Sterilization Efficiency: Pre-vacuum cycle removes air from the chamber, eliminating cold spots for more effective sterilization of solids
-
Multiple Application Modes: Specialized cycles for liquids (media/agar), solid materials, and waste/glassware
-
Real-Time Monitoring: Process display shows program progress and sterilization cycle status
-
Comprehensive Safety System: Lid safety interlock prevents opening during cycles, when pressurized, or above 97°C
-
Contamination Prevention: Bacteria filter reduces risk of re-contamination during cooling
-
User-Friendly Design: One-touch lever for easy lid operation with door-closure sensor
-
Space-Efficient Installation: Compact footprint with upward-opening lid minimizes required laboratory space
-
Mobile Configuration: Self-contained design mounted on casters allows flexible positioning
Laboratory Capacity Reference
- Can accommodate:
- 10L fermenter or 15L carboy
- 19" × 24" biohazard bag
- Six 1L flasks
- Twenty-one 500mL bottles
- 180 × 20mm test tubes (two stacked baskets)
- 270 × 13mm test tubes (three stacked baskets)
Package Includes
- Power cord
- Operator's manual
- 1-year manufacturer's warranty
The HICLAVE HVP-50 combines advanced vacuum technology with precise sterilization control to deliver superior drying performance and sterilization efficiency—ideal for laboratories requiring thorough drying of glassware, pipette tips, and other solid materials.
Specifications
Operating Parameters
Specification |
Details |
Temperature Operating Range |
105°C to 135°C |
Temperature Display Range |
5°C to 140°C |
Pressure Operating Range |
3 to 28.4 PSI (0.2 to 2.0 kg/cm²) |
Sterilization Timer |
1 to 250 minutes |
Drying Timer |
1 to 90 minutes |
Physical Specifications
Specification |
Details |
Chamber Volume |
50L |
Chamber Dimensions |
11.8" × 28" (30 × 71 cm) |
Overall Dimensions (W × D × H) |
20.1" × 24.8" × 40.9" (51 × 63 × 104 cm) |
Shipping Weight |
291 lbs (133 kg) |
Electrical Requirements
Specification |
Details |
Power Source |
120V, 50/60 Hz, single phase (220V/240V available on request) |
Power Consumption |
2000 watts (2000 VA) |
Program Cycles
Program Type |
Cycle Sequence |
Solids with Drying |
1. Pre-vacuum heating and air removal<br>2. Sterilization<br>3. Fast water and steam exhaust<br>4. Heat and vacuum drying<br>5. Completion |
Wastes/Solids without Drying |
1. Heating<br>2. Pre-vacuum<br>3. Sterilization<br>4. Fast steam exhaust<br>5. Completion |
Liquids/Agar |
1. Heating<br>2. Sterilization<br>3. Programmable auto steam exhaust (or ambient cooling without steam exhaust)<br>4. Completion |
Drying Only |
1. Heat and vacuum drying (programmable time)<br>2. Completion<br>For additional drying after Program #1 if desired
|
Safety Features
Feature |
Function |
Lid Interlock System |
Temperature- and pressure-sensitive mechanism prevents opening during operation |
Power Protection |
Over-pressure cutoff, over-temperature cutoff, low-water cutoff |
Emergency Relief |
Mechanical pressure relief valve |
Operation Safeguards |
Lid-closure detector, over-current detector |
Electrical Safety |
Circuit breaker with ground fault circuit interrupter (GFCI) |
Technical Specification Sheet
Please view the Specification sheet here